CLOSE

Micro - Balance XE Series
Micro - Balance XE Series
Micro - Balance XE Series
Micro - Balance XE Series
Micro - Balance XE Series
Micro - Balance XE Series
1/ 2

Micro - Balance XE Series

* Adopting a brand new supercell sensor technology - precise, fast, stable, and trustworthy. * Fully compliant with 21CFR Part 11- Full Traceability. * Adopting a full touchscreen operating system. * Adopting fully automatic calibration and adjustment technology.
Quantity
Only 10 left
Volume Pricing
Quantity
Price
1+
$20,000.00 USD each

Adopting a brand new supercell sensor technology - precise, fast, stable, and trustworthy

  • Sensitivity: Unparalleled sensor design and processing technology ensure the sensitivity of micro weighing results, up to a maximum of 0.1 μ g.
  • Precision: The unique super monomer sensor technology ensures the accuracy, stability, and consistency of micro weighing results.
  • Fast: Adopting the sensor technology of a horizontal bar, combined with excellent circuit design and software algorithms, it meets the requirements of laboratory symmetry measurement speed.

 

Fully compliant with 21CFR Part 11- Full Traceability

  • Designed in accordance with the data integrity principles of the US FDA, achieving accurate, clear, easy to read, synchronized, original, and traceable (ALCOA) data.
  • User management: Implement three-level user management through access control.
  • Electronic signature: Electronic signatures can be used to sign the final report of the weighing process.
  • Audit trail and weighing log: Audit trail is a time stamped, tamper proof electronic log file that allows for the reconstruction and recording of events related to creation, modification, and deletion.
  • Data transmission security: Electronic records should be protected from tampering.
  • The data output meets the requirements of GLP/GMP.

 

Adopting a full touchscreen operating system

  • The operation requirements are concise and clear, meeting the convenience of users, providing a good visual experience, while considering the special requirements of users in special industries, such as contactless operation, which can obtain accurate weighing results at any time and any temperature.

Adopting fully automatic calibration and adjustment technology

  • The IsoCAL (internal calibration and adjustment) function ensures accurate weighing results, using fully automatic temperature and time control calibration and adjustment functions.

 

Communication interface:

  • Standard communication interface, automatic recognition of printer.
  • Standard RS232, easy to connect with external devices.

 

Norms, laws, industry standards

  • GB/T26497-2022 Electronic balance
  • JB/T9511-1999 Technical Conditions for Balance Packaging
  • OIML R76-2 (2007E) EU: International Non Automatic Weighing Instruments

 

Specification

Model

XE2.6S

XE5.6S

XE31.2.6P

XE31.5.6P

XE31.6S

Capacity [Max]

2.1g

5.1g

2.1g/31g

5.1g/31g

31g

Readability [d]

1μg

1μg

1μg/10μg

1μg/10μg

1μg

Standard deviation

2μg

2μg/10μg

2μg

Typical error

2.5μg

2.5μg/20μg

2.5μg

Standard deviation

1.5μg

2.5μg

2.5μg/15μg

2.5μg

Typical error

3μg

4μg

3μg/20μg

5μg

Standard deviation

2μg

2μg/10μg

2μg

Typical error

2.5μg

2.5μg/20μg

5μg

Test weights

1g

2g

2g/20g

20g

Standard deviation

2μg

2μg

2μg/10μg

2μg

Typical error

2.5μg

2.5μg

2.5μg/20μg

5μg

Sensitivity drift (+10 ~ +30℃)

 

Accuracy level

 

Calibration division value (e)

(According to the 2014/31/EU standard)

10μg

10μg

10μg/100μg

10μg

Minimum weighing value

(According to the 2014/31/EU standard)

1mg

1mg

1mg

1mg

1mg

Min. optimal weighing

0.82mg

0.82mg

0.82mg

0.82mg

0.82mg

Min. typical weighing

1.3mg

1.3mg

1.3mg

1.3mg

1.3mg

Typical stability time

6s

6s

6s

6s

8s

Typical measurement time

8s

8s

8s

8s

12s

External calibration weights

2g

5g

2g/20g

5g/20g

20g

Standard weight grade

(According to OIML R111-1)

E1

E1

E1

E1

E1

Dimension

90×295×150 mm

Weighing Pan Dimensions

Φ30 mm

High of weighing chamber

80 mm

Net weight

4.7 kg

Operating temperature

18 ~ 23℃

Relative humidity

50% ~ 75%RH

Temperature compensation fluctuation

≤1°C/h

Power supply

AC100-240V, 50/60Hz, 1.0-0.5A